Medical Devices:

Federal Register Volume 76, Number 115 (Wednesday, June 15, 2011)

Rules and Regulations

Pages 34845-34847

From the Federal Register Online via the Government Printing Office [www.gpo.gov]

FR Doc No: 2011-14790

Federal Register

Rules and Regulations

Page 34845

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration 21 CFR Part 874

Docket No. FDA-2011-N-0361

Medical Devices; Ear, Nose, and Throat Devices; Classification of the Wireless Air-Conduction Hearing Aid

AGENCY: Food and Drug Administration, HHS.

ACTION: Final rule.

SUMMARY: The Food and Drug Administration (FDA) is classifying the wireless air-conduction hearing aid into class II (special controls).

The Agency is classifying the device into class II (special controls) in order to provide a reasonable assurance of safety and effectiveness of the device.

DATES: This rule is effective July 15, 2011. The classification was effective on March 31, 2011.

FOR FURTHER INFORMATION CONTACT: Vasant Dasika, Center for Devices and

Radiological Health, Food and Drug Administration, 10903 New Hampshire

Ave., Bldg. 66, rm. 2443, Silver Spring MD 20993-0002, 301-796-5365.

SUPPLEMENTARY INFORMATION:

  1. Background

    In accordance with section 513(f)(1) of the Federal Food, Drug, and

    Cosmetic Act (FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were not in commercial distribution before May 28, 1976 (the date of enactment of the Medical Device Amendments of 1976), generally referred to as postamendments devices, are classified automatically by statute into class III without any FDA rulemaking process. These devices remain in class III and require premarket approval, unless and until the device is classified or reclassified into class I or II, or FDA issues an order finding the device to be substantially equivalent, in accordance with section 513(i) of the FD&C Act, to a predicate device that does not require premarket approval. The Agency determines whether new devices are substantially equivalent to predicate devices by means of premarket notification procedures in section 510(k) of the FD&C Act (21

    U.S.C. 360(k)) and part 807 of the regulations (21 CFR part 807).

    Section 513(f)(2) of the FD&C Act provides that any person who submits a premarket notification under section 510(k) of the FD&C Act for a device that has not previously been classified may, within 30 days after receiving an order classifying the device into class III under section 513(f)(1), request FDA to classify the device under the criteria set forth in section 513(a)(1). FDA will, within 60 days of receiving this request, classify the device by written order. This classification will be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing this classification.

    In accordance with section 513(f)(1) of the FD&C Act, FDA issued an order on September 13, 2010, classifying the CLEAR 440 Series of hearing aids into class III, because it was not substantially equivalent to a device that was introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, or a device that was subsequently reclassified into class I or class II. On October 13, 2010, Widex Hearing Aid Co. submitted a petition requesting classification of the CLEAR 440 Series of hearing aids under section 513(f)(2) of the FD&C Act. The manufacturer recommended that the device be classified into class II

    (Ref. 1).

    In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed the petition in order to classify the device under the criteria for classification set forth in section 513(a)(1). FDA classifies devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the petition and information submitted during interactive review, FDA determined that the device can be classified into class II with the establishment of special controls. FDA believes these special controls will provide reasonable assurance of the safety and effectiveness of the device.

    The device is assigned the generic name wireless air-conduction hearing aid, and it is identified as a wearable sound-amplifying device, intended to compensate for impaired hearing, that incorporates wireless technology in its programming or use.

    FDA has identified the following risks to health associated with this type of device and the measures required to mitigate these risks:

    Identified risk

    Required measures

    Degradations in device function due to

    Electromagnetic compatibility electromagnetic interference (EMI).

    (EMC) testing; labeling.

    Degradations in device function due to

    Wireless technology design, wireless technology disruption such as description, and testing; slow-down, lost or corrupted

    performance testing; information, security issues including labeling. potential cross-talk or control by other users with a similar medical device.

    Exposure to non-ionizing radiation

    Wireless technology design, emitted by wireless technology can

    description, analysis, and potentially induce tissue heating.

    testing; labeling.

    FDA believes that the following special controls, in addition to general controls, address the risks to health and provide reasonable assurance of the safety and effectiveness of the device: (1)

    Appropriate analysis/testing should validate EMC and safety of exposure to non-ionizing radiation; (2) Design, description, and performance data

    Page 34846

    should validate wireless technology functions; and (3) Labeling should specify appropriate instructions, warnings, and information relating to

    EMC and wireless technology and human exposure to non-ionizing radiation. Therefore, on March 31, 2011, FDA issued an order to the petitioner classifying the device into class II. FDA is codifying the classification of the device by adding Sec. 874.3305.

    Following the effective date of this final classification rule, any firm introducing a wireless air-conduction hearing aid into interstate commerce in the United States will need to comply with the special controls named in the regulation. However, the firm need only show that its device meets the recommendations of the special controls or in some other way provides equivalent assurance of safety and effectiveness.

    Section 510(m) of the FD&C Act provides that FDA may exempt a class

    II device from the premarket notification requirements under section 510(k) if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. For this type of device, FDA has determined that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device. Therefore, this device type is exempt from premarket notification requirements.

  2. Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

  3. Analysis of Impacts

    FDA has examined the impacts of the final rule under Executive

    Order 12866, Executive Order 13563, the Regulatory Flexibility Act (5

    U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Pub. L. 104-4). Executive Order 12866 directs Agencies to assess all costs and benefits of available regulatory alternatives and, when regulation is necessary, to select regulatory approaches that maximize net benefits

    (including potential economic, environmental, public health and safety, and other advantages; distributive impacts; and equity). The Agency believes that this final rule is not a significant regulatory action under Executive Order 12866.

    The Regulatory Flexibility Act requires Agencies to analyze regulatory options that would minimize any significant impact of a rule on small entities. Because reclassification of this device from class

    III to class II will relieve manufacturers of the device of the cost of complying with the premarket approval requirements of section 515 of the FD&C Act (21 U.S.C. 360e), and may permit small potential competitors to enter the marketplace by lowering their costs, the

    Agency certifies that the final rule will not have a significant economic impact on a substantial number of small entities.

    Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires that Agencies prepare a written statement, which includes an assessment of anticipated costs and benefits, before proposing ``any rule that includes any Federal mandate that may result in the expenditure by

    State, local, and Tribal governments, in the aggregate, or by the private sector, of $100,000,000 or more (adjusted annually for inflation) in any one year.'' The current threshold after adjustment for inflation is $136 million, using the most current (2010) Implicit

    Price Deflator for the Gross Domestic Product. FDA does not expect this final rule to result in any 1-year expenditure that would meet or exceed this amount.

  4. Federalism

    FDA has analyzed this final rule in accordance with the principles set forth in Executive Order 13132. Section 4(a) of the Executive order requires Agencies to ``construe * * * a Federal statute to preempt

    State law only where the statute contains an express preemption provision or there is some other clear evidence that the Congress intended preemption of State law, or where the exercise of State authority conflicts with the exercise of Federal authority under the

    Federal statute.'' Federal law includes an express preemption provision that preempts certain state requirements ``different from or in addition to'' certain Federal requirements applicable to devices. (See section 521 of the FD&C Act (21 U.S.C. 360k); See Medtronic v. Lohr, 518 U.S. 470 (1996); Riegel v. Medtronic, Inc., 552 U.S. 312 (2008).

    The special controls established by this final rule create

    ``requirements'' for specific medical devices under 21 U.S.C. 360k, even though product sponsors have some flexibility in how they meet those requirements. (See Papike v. Tambrands, Inc., 107., 107 F.3d 737, 740-42 (9th Cir. 1997).)

  5. Paperwork Reduction Act of 1995

    FDA concludes that this final rule contains no new collections of information. Therefore, clearance by the Office of Management and

    Budget (OMB) under the Paperwork Reduction Act of 1995 (PRA) (44 U.S.C. 3501-3520) is not required.

  6. References

    The following references have been placed on display in the

    Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852, and may be seen by interested persons between 9 a.m. and 4 p.m., Monday through Friday. 1. Petition from Widex Hearing Aid Co., dated October 13, 2010.

    List of Subjects in 21 CFR Part 874

    Medical devices.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 874 is amended as follows:

    PART 874--EAR, NOSE, AND THROAT DEVICES 0 1. The authority citation for 21 CFR part 874 continues to read as follows:

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371. 0 2. Section Sec. 874.3305 is added to subpart D to read as follows:

    Sec. 874.3305 Wireless Air-Conduction Hearing Aid.

    (a) Identification. A wireless air-conduction hearing aid is a wearable sound-amplifying device, intended to compensate for impaired hearing that incorporates wireless technology in its programming or use.

    (b) Classification: Class II (special controls). The special controls for this device are:

    (1) Appropriate analysis/testing should validate electro magnetic compatibility (EMC) and safety of exposure to non-ionizing radiation;

    (2) Design, description, and performance data should validate wireless technology functions; and

    (3) Labeling should specify appropriate instructions, warnings, and information relating to EMC and wireless technology and human exposure to non-ionizing radiation.

    (c) Premarket notification. The wireless air-conduction hearing aid is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to Sec. 874.9.

    Page 34847

    Dated: June 9, 2011.

    Nancy K. Stade,

    Deputy Director for Policy, Center for Devices and Radiological Health.

    FR Doc. 2011-14790 Filed 6-14-11; 8:45 am

    BILLING CODE 4160-01-P