Medical Devices; Gastroenterology-Urology Devices; Classification of the Prostatic Artery Embolization Device

Federal Register, Volume 82 Issue 218 (Tuesday, November 14, 2017)

Federal Register Volume 82, Number 218 (Tuesday, November 14, 2017)

Rules and Regulations

Pages 52649-52651

From the Federal Register Online via the Government Publishing Office www.gpo.gov

FR Doc No: 2017-24586

-----------------------------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Part 876

Docket No. FDA-2017-N-6289

Medical Devices; Gastroenterology-Urology Devices; Classification of the Prostatic Artery Embolization Device

AGENCY: Food and Drug Administration, HHS.

ACTION: Final order.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA or we) is classifying the prostatic artery embolization device into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the prostatic artery embolization device's classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

DATES: This order is effective November 14, 2017. The classification was applicable on June 21, 2017.

FOR FURTHER INFORMATION CONTACT: Benjamin Fisher, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. G108, Silver Spring, MD 20993-0002, 301-

796-0245, Benjamin.Fisher@fda.hhs.gov.

SUPPLEMENTARY INFORMATION:

  1. Background

    Upon request, FDA has classified the prostatic artery embolization device as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as ``postamendments devices'' because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act to a predicate device that does not require premarket approval (see 21 U.S.C. 360c(i)). We determine whether a new device is substantially equivalent to a predicate by means of the procedures

    Page 52650

    for premarket notification under section 510(k) of the FD&C Act and part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).

    FDA may also classify a device through ``De Novo'' classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act (21 U.S.C. 360c(f)(2)). Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-

    144). A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA shall classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically within class III, the De Novo classification is considered to be the initial classification of the device.

    We believe this De Novo classification will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens. When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a result, other device sponsors do not have to submit a De Novo request or pre-market approval in order to market a substantially equivalent device (see 21 U.S.C. 360c(i), defining ``substantial equivalence''). Instead, sponsors can use the less-burdensome 510(k) process, when necessary, to market their device.

  2. De Novo Classification

    On August 5, 2016, BioSphere Medical, S.A., submitted a request for De Novo classification of the Embospheresupreg Microspheres. FDA reviewed the request to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act.

    We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the generals controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to the general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Therefore, on June 21, 2017, FDA issued an order to the requester classifying the device into class II. FDA is codifying the classification of the device by adding 21 CFR 876.5550. We have named the generic type of device prostatic artery embolization device, and it is identified as an intravascular implant intended to occlude the prostatic arteries to prevent blood flow to the targeted area of the prostate, resulting in a reduction of lower urinary tract symptoms related to benign prostatic hyperplasia. This does not include cyanoacrylates and other embolic agents which act by in situ polymerization or precipitation, or embolization devices used in neurovascular applications (see 21 CFR 882.5950).

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1--Prostatic Artery Embolization Device Risks and Mitigation

    Measures

    ------------------------------------------------------------------------

    Identified risks Mitigation measures

    ------------------------------------------------------------------------

    Adverse tissue reaction................ Biocompatibility evaluation.

    Infection.............................. Sterilization validation, Shelf-

    life validation, Non-clinical

    performance testing, and

    Labeling.

    Non-target ischemia.................... Clinical data, Non-clinical

    performance testing, and

    Labeling.

    Urinary retention...................... Labeling.

    Post-prostatic artery embolization Labeling.

    syndrome (nausea, vomiting, regional

    pain, non-infectious fever, minor

    hematuria, or hematochezia).

    ------------------------------------------------------------------------

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. In order for a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. This device is subject to premarket notification requirements under section 510(k).

  3. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

  4. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in the guidance document ``De Novo Classification Process (Evaluation of Automatic Class III Designation)'' have been approved under OMB control number 0910-0844; the collections of information in part 21 CFR part 814, subparts A through E, regarding premarket approval, have been approved under OMB control number 0910-0231; the collections of

    Page 52651

    information in part 807, subpart E, regarding premarket notification submissions, have been approved under OMB control number 0910-0120; and the collections of information in 21 CFR part 801, regarding labeling, have been approved under OMB control number 0910-0485.

    List of Subjects in 21 CFR Part 876

    Medical devices.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 876 is amended as follows:

    PART 876--GASTROENTEROLOGY-UROLOGY DEVICES

    0

    1. The authority citation for part 876 continues to read as follows:

      Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

      0

    2. Add Sec. 876.5550 to subpart F to read as follows:

      Sec. 876.5550 Prostatic artery embolization device.

      (a) Identification. A prostatic artery embolization device is an intravascular implant intended to occlude the prostatic arteries to prevent blood flow to the targeted area of the prostate, resulting in a reduction of lower urinary tract symptoms related to benign prostatic hyperplasia. This does not include cyanoacrylates and other embolic agents which act by in situ polymerization or precipitation, or embolization devices used in neurovascular applications (see 21 CFR 882.5950).

      (b) Classification. Class II (special controls). The special controls for this device are:

      (1) The device must be demonstrated to be biocompatible.

      (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

      (i) Evaluation of suitability for injection through catheters intended for use in embolization; and

      (ii) Evaluation of the size distribution of the device.

      (3) Performance data must support the sterility and pyrogenicity of the device.

      (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.

      (5) Clinical data must evaluate post-embolization damage due to non-target embolization under anticipated use conditions.

      (6) The labeling must include:

      (i) Specific instructions on safe device preparation and use;

      (ii) The device shelf life;

      (iii) Data regarding urinary retention; and

      (iv) Data regarding post-prostatic artery embolization syndrome.

      Dated: November 7, 2017.

      Lauren Silvis,

      Chief of Staff.

      FR Doc. 2017-24586 Filed 11-13-17; 8:45 am

      BILLING CODE 4164-01-P