Medical Devices; Immunology and Microbiology Devices; Classification of the Mass Spectrometer System for Clinical Use for the Identification of Microorganisms

Federal Register, Volume 82 Issue 204 (Tuesday, October 24, 2017)

Federal Register Volume 82, Number 204 (Tuesday, October 24, 2017)

Rules and Regulations

Pages 49100-49102

From the Federal Register Online via the Government Publishing Office www.gpo.gov

FR Doc No: 2017-23022

-----------------------------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Part 866

Docket No. FDA-2017-N-5290

Medical Devices; Immunology and Microbiology Devices; Classification of the Mass Spectrometer System for Clinical Use for the Identification of Microorganisms

AGENCY: Food and Drug Administration, HHS.

ACTION: Final order.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA or we) is classifying the mass spectrometer system for clinical use for the identification of microorganisms into class II (special controls). The special controls that apply to the device type are identified in this order and will be part of the codified language for the mass spectrometer system for clinical use for the identification of microorganisms' classification. We are taking this action because we have determined that classifying the device into class II (special controls) will provide a reasonable assurance of safety and effectiveness of the device. We believe this action will also enhance patients' access to beneficial innovative devices, in part by reducing regulatory burdens.

DATES: This order is effective October 24, 2017. The classification was applicable on August 21, 2013.

FOR FURTHER INFORMATION CONTACT: Steven Tjoe, Center for Devices and Radiological Health, Food and Drug Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 4550, Silver Spring, MD 20993-0002, 301-796-5866, steven.tjoe@fda.hhs.gov.

SUPPLEMENTARY INFORMATION:

  1. Background

    Upon request, FDA has classified the mass spectrometer system for clinical use for the identification of microorganisms as class II (special controls), which we have determined will provide a reasonable assurance of safety and effectiveness. In addition, we believe this action will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens by placing the device into a lower device class than the automatic class III assignment.

    The automatic assignment of class III occurs by operation of law and without any action by FDA, regardless of the level of risk posed by the new device. Any device that was not in commercial distribution before May 28, 1976, is automatically classified as, and remains within, class III and requires premarket approval unless and until FDA takes an action to classify or reclassify the device (see 21 U.S.C. 360c(f)(1)). We refer to these devices as ``postamendments devices'' because they were not in commercial distribution prior to the date of enactment of the Medical Device Amendments of 1976, which amended the Federal Food, Drug, and Cosmetic Act (the FD&C Act).

    FDA may take a variety of actions in appropriate circumstances to classify or reclassify a device into class I or II. We may issue an order finding a new device to be substantially equivalent under section 513(i) of the FD&C Act to a predicate device that does not require premarket approval (see 21 U.S.C. 360c(i)). We determine whether a new device is substantially equivalent to a predicate by means of the procedures for premarket notification under section 510(k) of the FD&C Act and part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).

    FDA may also classify a device through ``De Novo'' classification, a common name for the process authorized under section 513(f)(2) of the FD&C Act. Section 207 of the Food and Drug Administration Modernization Act of 1997 established the first procedure for De Novo classification (Pub. L. 105-115). Section 607 of the Food and Drug Administration Safety and Innovation Act modified the De Novo application process by adding a second procedure (Pub. L. 112-144). A device sponsor may utilize either procedure for De Novo classification.

    Under the first procedure, the person submits a 510(k) for a device that has not previously been classified. After receiving an order from FDA classifying the device into class III under section 513(f)(1) of the FD&C Act, the person then requests a classification under section 513(f)(2).

    Under the second procedure, rather than first submitting a 510(k) and then a request for classification, if the person determines that there is no legally marketed device upon which to base a determination of substantial equivalence, that person requests a classification under section 513(f)(2) of the FD&C Act.

    Under either procedure for De Novo classification, FDA shall classify the device by written order within 120 days. The classification will be according to the criteria under section 513(a)(1) of the FD&C Act. Although the device was automatically placed within class III, the De Novo classification is considered to be the initial classification of the device.

    We believe this De Novo classification will enhance patients' access to beneficial innovation, in part by reducing regulatory burdens. When FDA classifies a device into class I or II via the De Novo process, the device can serve as a predicate for future devices of that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a result, other device sponsors do not have to submit a De Novo request or premarket approval application (PMA) in order to market a substantially equivalent device (see 21 U.S.C. 360c(i), defining ``substantial equivalence''). Instead, sponsors can use the less-

    burdensome 510(k) process, when necessary, to market their device.

  2. De Novo Classification

    On January 3, 2013, bioMeacuterieux, Inc. submitted a request for De Novo classification of the Viteksupreg MS. FDA reviewed the request in order to classify the device under the criteria for classification set forth in section 513(a)(1) of the FD&C Act. We classify devices into class II if general controls by themselves are insufficient to provide reasonable assurance of safety and effectiveness, but there is sufficient information to establish special controls that, in combination with the general controls, provide reasonable assurance of the safety and effectiveness of the device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the information submitted in the request, we determined that the device can be classified into class II with the establishment of special controls. FDA has determined that these special controls, in addition to general controls, will provide reasonable assurance of the safety and effectiveness of the device.

    Page 49101

    Therefore, on August 21, 2013, FDA issued an order to the requestor classifying the device into class II. FDA is codifying the classification of the device by adding 21 CFR 866.3361. We have named the generic type of device mass spectrometer system for clinical use for the identification of microorganisms, and it is identified as a qualitative in vitro diagnostic device intended for the identification of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer, and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.

    FDA has identified the following risks to health associated specifically with this type of device and the measures required to mitigate these risks in table 1.

    Table 1--Mass Spectrometer System for Clinical Use for the

    Identification of Microorganisms Risks and Mitigation Measures

    ------------------------------------------------------------------------

    Identified risks Mitigation measures

    ------------------------------------------------------------------------

    Incorrect identification of a (1) Premarket notification

    pathogenic microorganism can lead to submissions must include

    improper patient management. detailed documentation for

    device software, including,

    but not limited to, standalone

    software applications and

    hardware-based devices that

    incorporate software.

    (2) Premarket notification

    submissions must include

    database implementation

    methodology, construction

    parameters, and quality

    assurance protocols.

    Failure to correctly interpret test (1) A detailed explanation of

    results. the interpretation of results

    and acceptance criteria must

    be included in the device's 21

    CFR 809.10(b)(9) compliant

    labeling.

    Failure to correctly operate the (1) As part of the risk

    instrument. management activities

    performed as part of your 21

    CFR 820.30 design controls,

    you must document an

    appropriate end user device

    training program that will be

    offered as part of your

    efforts to mitigate the risk

    of failure to correctly

    operate the instrument.

    (2) Premarket notification

    submissions must include

    details on the appropriate end

    user device training program

    that will be offered while

    marketing the device.

    ------------------------------------------------------------------------

    FDA has determined that special controls, in combination with the general controls, address these risks to health and provide reasonable assurance of safety and effectiveness. In order for a device to fall within this classification, and thus avoid automatic classification in class III, it would have to comply with the special controls named in this final order. The necessary special controls appear in the regulation codified by this order. The device is subject to premarket notification requirements under section 510(k).

  3. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is of a type that does not individually or cumulatively have a significant effect on the human environment. Therefore, neither an environmental assessment nor an environmental impact statement is required.

  4. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to previously approved collections of information found in other FDA regulations. These collections of information are subject to review by the Office of Management and Budget (OMB) under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of information in part 807, subpart E, regarding premarket notification submissions have been approved under OMB control number 0910-0120, the collections of information in 21 CFR part 820 have been approved under OMB control number 0910-0073, and the collections of information in 21 CFR parts 801 and 809, regarding labeling have been approved under OMB control number 0910-0485.

    List of Subjects in 21 CFR Part 866

    Biologics, Laboratories, Medical devices.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, 21 CFR part 866 is amended as follows:

    PART 866--IMMUNOLOGY AND MICROBIOLOGY DEVICES

    0

    1. The authority citation for part 866 continues to read as follows:

      Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.

      0

    2. Add Sec. 866.3361 to subpart D to read as follows:

      Sec. 866.3361 Mass spectrometer system for clinical use for the identification of microorganisms.

      (a) Identification. A mass spectrometer system for clinical use for the identification of microorganisms is a qualitative in vitro diagnostic device intended for the identification of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer, and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.

      (b) Classification. Class II (special controls). The special controls for this device are:

      (1) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.

      (2) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols.

      (3) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.

      (4) As part of the risk management activities performed as part of your 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.

      Page 49102

      (5) Premarket notification submissions must include details on the appropriate end user device training program that will be offered while marketing the device.

      Dated: October 19, 2017.

      Leslie Kux,

      Associate Commissioner for Policy.

      FR Doc. 2017-23022 Filed 10-23-17; 8:45 am