General and Plastic Surgery Devices: Proposed Classification for the Tissue Expander

Federal Register: December 22, 2008 (Volume 73, Number 246)

Proposed Rules

Page 78239-78242

From the Federal Register Online via GPO Access [wais.access.gpo.gov]

DOCID:fr22de08-24

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration 21 CFR Part 878

Docket No. FDA-2008-N-0604

General and Plastic Surgery Devices: Proposed Classification for the Tissue Expander

AGENCY: Food and Drug Administration, HHS.

ACTION: Proposed rule.

SUMMARY: The Food and Drug Administration (FDA) is proposing to classify into class II (special controls) the tissue expander, as a device intended for temporary (less than 6 months) subdermal implantation to stretch the skin for surgical applications, specifically to develop surgical flaps and additional tissue coverage.

Elsewhere in this issue of the Federal Register, FDA is announcing the availability of the draft guidance that FDA intends will serve as the special control if FDA classifies this device type into class II.

DATES: Submit written or electronic comments by March 23, 2009. See section IV of this document for the proposed effective date of a final rule based on this proposed rule.

ADDRESSES: You may submit comments, identified by Docket No. FDA-2008-

N-0604, by any of the following methods:

Electronic Submissions

Submit electronic comments in the following way:

Federal eRulemaking Portal: http://www.regulations.gov.

Follow the instructions for submitting comments.

Written Submissions

Submit written submissions in the following ways:

FAX: 301-827-6870.

Mail/Hand delivery/Courier [For paper, disk, or CD-ROM submissions]: Division of Dockets Management (HFA-305), Food and Drug

Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852.

To ensure more timely processing of comments, FDA is no longer accepting comments submitted to the agency by e-mail. FDA encourages you to continue to submit electronic comments by using the Federal eRulemaking Portal, as described previously, in the ADDRESSES portion of this document under Electronic Submissions.

Instructions: All submissions received must include the agency name and docket number for this rulemaking. All comments received may be posted without change to http://www.regulations.gov, including any personal information provided. For additional information on submitting comments, see the ``Comments'' heading of the SUPPLEMENTARY INFORMATION section of this document.

Docket: For access to the docket to read background documents or comments received, go to http://www.regulations.gov and insert the docket number, found in brackets in the heading of this document, into the ``Search'' box and follow the prompts and/or go to the Division of

Dockets Management, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852.

FOR FURTHER INFORMATION CONTACT: Nada Hanafi, Center for Devices and

Radiological Health (HFZ-4), Food and Drug Administration, 7520

Standish Pl., Rockville, MD 20855, 240-276-8848.

SUPPLEMENTARY INFORMATION:

  1. Background

    The Federal Food, Drug, and Cosmetic Act (the act), as amended by the Medical Device Amendments of 1976 (the 1976 amendments) (Public Law 94-295), the Safe Medical Devices Act of 1990 (Public Law 101-629), and the Food and Drug Modernization Act of 1997 (FDAMA) (Public Law 105- 115), the Food and Drug Administration Amendments Act of 2007 (Public

    Law 110-85), among other amendments, established a comprehensive system for the regulation of medical devices intended for human use. Section 513 of the act (21 U.S.C. 360c) established three categories (classes) of devices, depending on the regulatory controls needed to provide reasonable assurance of their safety and effectiveness. The three categories of devices are class I (general controls), class II (special controls), and class III (premarket approval).

    Under section 513 of the act, FDA refers to devices that were in commercial distribution before May 28, 1976 (the date of enactment of the 1976 amendments), as ``preamendments devices.'' FDA classifies these devices after the agency has taken the following steps:

    Page 78240

    (1) Receives a recommendation from a device classification panel

    (an FDA advisory committee);

    (2) Publishes the panel's recommendation for comment, along with a proposed regulation classifying the device type; and

    (3) Publishes a final regulation classifying the device type.

    FDA has classified most preamendments devices under these procedures.

    FDA refers to devices that were not in commercial distribution before May 28, 1976, as ``postamendments devices.'' These device types are classified automatically by statute (section 513(f) of the act) into class III without any FDA rulemaking process. Those device types remain in class III and require premarket approval, unless and until:

    (1) FDA reclassifies the device type into class I or II;

    (2) FDA issues an order classifying the device type into class I or

    II in accordance with new section 513(f)(2) of the act, as amended by

    FDAMA; or

    (3) FDA issues an order finding the device to be substantially equivalent, under section 513(i) of the act, to a predicate device that does not require premarket approval.

    The agency determines whether new devices are substantially equivalent to previously offered devices by means of premarket notification procedures in section 510(k) of the act (21 U.S.C. 360(k)) and 21 CFR part 807 of the regulations.

    A person may market a preamendments device that has been classified into class III through premarket notification procedures, without submission of a premarket approval application (PMA), until FDA issues a final regulation under section 515(b) of the act (21 U.S.C. 360e(b)) requiring premarket approval.

    The tissue expander is a preamendment device type that was not classified in the final rule published in the Federal Register of June 24, 1988, classifying other General and Plastic Surgery Devices (53 FR 23856). Consistent with the act and the regulations, FDA consulted with the Panel, an FDA advisory committee, regarding the classification of this device type.

  2. Recommendation of the Panel

    At a public meeting held on August 25 and 26, 2005, the Panel unanimously recommended that the tissue expander be classified into class II (Ref. 1). The Panel believed that class II, special controls, in addition to general controls, would reasonably assure the safety and effectiveness of this device type. The Panel also recommended that the special control for the device type be a guidance document.

    1. Identification

      FDA is proposing the following identification based on the Panel's recommendation and the available information: A tissue expander is a device intended for temporary (less than 6 months) subdermal implantation to stretch the skin for surgical applications, specifically to develop surgical flaps and additional tissue coverage.

      It is made of an inflatable silicone elastomer shell filled with Normal

      Physiological Saline (injection grade).

    2. Recommended Classification of the Panel

      The Panel unanimously recommended that the tissue expander be classified into class II. The Panel believed that class II with the special controls (a guidance document and labeling) would provide reasonable assurance of the safety and effectiveness of the device.

      Elsewhere in this issue of the Federal Register, FDA is announcing the availability of the draft guidance that will serve as the special control for this device type.

    3. Summary of Reasons for Recommendation

      After reviewing the information provided by FDA, and after consideration of the open discussions during the Panel meeting and the

      Panel members' personal knowledge of and clinical experience with the device system, the Panel provided the following reasons in support of its recommendation to classify the generic device type, tissue expander intended for temporary (less than 6 months) subdermal implantation to develop surgical flaps and additional coverage for surgical applications, into class II. The Panel believed the tissue expander should be classified into class II because special controls, in addition to general controls, would provide reasonable assurance of the safety and effectiveness of the device and there is sufficient information to establish special controls to provide such assurance.

    4. Summary of the Data Upon Which the Recommendation is Based

      In addition to the potential risks to health associated with implantation of the tissue expander described in section II.E of this document, ``Risks to Health,'' there is reasonable knowledge of the benefits of the device type. Specifically, the tissue expander develops tissue flaps and coverage needed for surgical applications, such as breast reconstruction following mastectomy, treatment of underdeveloped breasts, scar revision, and treatment of soft tissue deformities or injuries.

    5. Risks to Health

      After considering the Panel's comments and recommendation, the published literature, and medical device reports, FDA has evaluated the risks to health associated with use of the tissue expander. FDA believes the following are risks to health associated with use of the device type:

      Skin trauma, including necrosis, thinning and slough;

      Device failure, including rupture and injection site/port failure;

      Infection--Infection is a risk to health associated with all surgical procedures and implanted devices. Incompatible or impure material composition may irritate the surrounding tissue which could increase the risk of infection. Use of a device that is not pyrogen free may elicit a fever.

      Adverse tissue reaction--Adverse tissue reaction is a risk to health common to all implanted devices. The implantation of the tissue expander will elicit a mild inflammatory reaction typical of a normal foreign body response. Incompatible material or impurities in the materials may increase the severity of a local tissue reaction or cause a systemic tissue reaction.

      Pain--Pain is a risk to health associated with all surgical procedures and implanted devices.

    6. Special Controls

      In addition to general controls, FDA believes that the draft guidance document entitled ``Class II Special Controls Guidance: Tissue

      Expander'' (the draft class II special controls guidance document) is a special control adequate to address the risks to health associated with the use of the device type described in section II.E of this document.

      FDA believes that the draft class II special controls guidance document addresses the Panel's concerns and provides reasonable assurance of the safety and effectiveness of the device type. Elsewhere in this issue of the Federal Register, FDA is publishing a notice of availability of the draft class II special controls guidance document that the agency would use as the special control for this device type.

      The draft class II special controls guidance document sets forth the information FDA recommends submitters include in premarket notification submissions (510(k)s) for a tissue expander. FDA has identified the risks to health associated with the use

      Page 78241

      of the device type in the first column of table 1 of this document. The recommended mitigation measures identified in the draft class II special controls guidance document is in the second column of table 1 of this document. FDA believes that addressing these risks to health in a 510(k) in the manner identified in the draft class II special controls guidance document, or in an acceptable alternative manner, is necessary to provide reasonable assurance of the safety and effectiveness of the device type.

      Table 1.--Risks to Health and Mitigation Measures

      Identified Risk

      Recommended Mitigation Measures

      Skin trauma (e.g., necrosis,

      Labeling thinning, sloughing).

      Device failure (e.g., rupture,

      Preclinical testing injection site/port failure).

      Labeling

      Infection............................ Sterility

      Adverse tissue reaction.............. Biocompatibility

      Pain................................. Labeling

  3. Proposed Classification

    FDA concurs with the Panel's recommendation that a tissue expander should be classified into class II because special controls, in addition to general controls, would provide reasonable assurance of the safety and effectiveness of the device, and there is sufficient information to establish special controls to provide such assurance.

  4. Proposed Effective Date

    FDA proposes that any final regulation based on this proposal become effective 30 days after its date of publication in the Federal

    Register.

  5. Environmental Impact

    The agency has determined under 21 CFR 25.34(b) that this proposed classification action is of a type that does not individually or cumulatively have a significant effect on the human environment.

    Therefore, neither an environmental assessment nor an environmental impact statement is required.

  6. Analysis of Impacts

    FDA has examined the impacts of the proposed rule under Executive

    Order 12866 and the Regulatory Flexibility Act (5 U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Public Law 104-4). Executive

    Order 12866 directs agencies to assess all costs and benefits of available regulatory alternatives and, when regulation is necessary, to select regulatory approaches that maximize net benefits (including potential economic, environmental, public health and safety, and other advantages; distributive impacts; and equity). The agency believes that this proposed rule is not a significant regulatory action as defined by the Executive order.

    The Regulatory Flexibility Act requires agencies to analyze regulatory options that would minimize any significant impact of a rule on small entities. Classification of this device type into class II will have a negligible impact on manufacturers because manufacturers of the device type currently must provide premarket notification before marketing the device and because FDA believes that manufacturers are already substantially in compliance with the recommendations in the draft guidance document. Because classification into class II will not increase regulatory costs with respect to this device type, the agency proposes to certify that the final rule will not have a significant economic impact on a substantial number of small entities.

    Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires that agencies prepare a written statement, which includes an assessment of anticipated costs and benefits, before proposing ``any rule that includes any Federal mandate that may result in the expenditure by

    State, local, and tribal governments, in the aggregate, or by the private sector, of $100,000,000 or more (adjusted annually for inflation) in any one year.'' The current threshold after adjustment for inflation is $130 million, using the most current (2007) Implicit

    Price Deflator for the Gross Domestic Product. FDA does not expect this proposed rule to result in any 1-year expenditure that would meet or exceed this amount.

  7. Federalism

    FDA has analyzed this proposed rule in accordance with the principles set forth in Executive Order 13132. Section 4(a) of the

    Executive order requires agencies to ``construe * * * a Federal statute to preempt State law only where the statute contains an express preemption provision or there is some other clear evidence that the

    Congress intended preemption of State law, or where the exercise of

    State authority conflicts with the exercise of Federal authority under the Federal statute.'' Federal law includes an express preemption provision that preempts certain State requirements ``different or in addition to'' certain federal requirements applicable to devices (21

    U.S.C. 360k; Medtronic v. Lohr, 518 U.S. 470 (1996); Riegel v.

    Medtronic, 128 S.Ct. 999 (2008)). In this proposed rulemaking, FDA has tentatively determined that general controls by themselves are insufficient to provide reasonable assurance of the safety and effectiveness of the device, and that there is sufficient information to establish special controls to provide such assurance. FDA therefore proposes to establish special controls to address the issues of safety or effectiveness identified in the special controls draft guidance document. If this proposed rule is made final, these special controls would create ``requirements'' for specific medical devices under 21

    U.S.C. 360k, even though product sponsors would have some flexibility in how they meet those requirements (Papike v. Tambrands, Inc., 107

    F.3d 737, 740-42 (9th Cir. 1997)).

    In addition, if this rule becomes final, as with any Federal requirement, if a State law requirement makes compliance with both

    Federal law and State law impossible, or would frustrate Federal objectives, the State requirement would be preempted. (See Geier v.

    American Honda Co., 529 U.S. 861 (2000); English v. General Electric

    Co., 496 U.S. 72, 79 (1990); Florida Lime & Avocado Growers, Inc., 373

    U.S. 132, 142-43 (1963); Hines v. Davidowitz, 312 U.S. 52, 67 (1941).)

  8. Paperwork Reduction Act of 1995

    FDA tentatively concludes that this proposed rule contains no new collections of information. Therefore, clearance by the Office of

    Management and Budget under the Paperwork Reduction Act of 1995 (44

    U.S.C. 3501-3520) is not required. This proposed rule designates a guidance document as a special control.

    Elsewhere in this issue of the Federal Register, FDA is publishing a notice of availability of the draft guidance document entitled

    ``Class II Special Controls Guidance Document: Tissue Expander,'' which contains an analysis of the paperwork burden for the draft guidance.

  9. Comments

    Interested persons may submit to the Division of Dockets Management

    (see ADDRESSES) written or electronic comments regarding this document.

    Submit a single copy of electronic comments or two paper copies of any mailed comments, except that individuals may submit one paper copy.

    Page 78242

    Comments are to be identified with the docket number found in brackets in the heading of this document. Received comments may be seen in the

    Division of Dockets Management between 9 a.m. and 4 p.m., Monday through Friday.

    Please note that on January 15, 2008, the FDA Division of Dockets

    Management Web site transitioned to the Federal Dockets Management

    System (FDMS). FDMS is a Government-wide, electronic docket management system. Electronic comments or submissions will be accepted by FDA only through FDMS at http://www.regulations.gov.

  10. References

    The following reference has been placed on display in the Division of Dockets Management (see ADDRESSES) and may be seen by interested persons between 9 a.m. and 4 p.m., Monday through Friday. 1. General and Plastic Surgery Devices Panel, Transcript, August 25 and 26, 2005, pp. 11 through 58 of the August 26, 2005, transcripts.

    List of Subjects in 21 CFR Part 878

    Medical devices.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under authority delegated to the Commissioner of Food and Drugs, FDA proposes to amend 21 CFR part 878 as follows:

    PART 878--GENERAL AND PLASTIC SURGERY DEVICES 1. The authority citation for 21 CFR part 878 continues to read as follows:

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 3601, 371. 2. Add Sec. 878.3600 to subpart D to read as follows:

    Sec. 878.3600 Tissue expander.

    (a) Identification. A tissue expander is a device intended for temporary (less than 6 months) subdermal implantation to stretch the skin for surgical applications, specifically to develop surgical flaps and additional tissue coverage. It is made of an inflatable silicone elastomer shell filled with Normal Physiological Saline (injection grade).

    (b) Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled ``Class II

    Special Controls Guidance Document: Tissue Expander.'' See Sec. 878.1(e) for availability information of guidance documents.

    Dated: December 16, 2008.

    Jeffrey Shuren,

    Associate Commissioner for Policy and Planning.

    FR Doc. E8-30439 Filed 12-19-08; 8:45 am